email HTI
MODEL T750
HIGH-VOLTAGE DRIVER
  • Four channel, individually transformer-isolated pulse driver

  • Adjustable pulse amplitude 5 volts to 100 volts into 50 ohm load

  • Outputs clean, flat pulses with typical rise/fall times below 2 ns

  • External universal power supply or 12-volt DC power.

  • Ideal companion to the Highland T560 embedded digital delay generator

  • OEM packaged or board-only custom available
Waveform shown: 50 volt pulse into 50 ohms
The T750 is a compact 4-channel high-voltage pulse driver. Each channel includes an input comparator and a transformer-coupled high-voltage power output stage. Output pulses follow the logic inputs and are processed by an adaptive pulse width limiter that prevents output transformer saturation.

Highland Technology also offers the T760, a 2-channel high-volrage driver with fiberoptic inputs

Specifications: T750 High-Voltage Driver

FUNCTION 4 Channel, General-purpose, Isolated High Voltage Pulse Driver
PACKAGING 4.0 x 4.75 x 1.25 inches extruded aluminum enclosure.
INPUTS Four single-ended, positive pulse inputs

+5 Volts max

50 ohm terminations

+1. 0 V minimum usable input

Common threshold trimpot, comparator level +0.5 to +2.5V. Threshold-set test point is provided.

OUTPUT Four individually-isolated transformer-coupled positive pulses

Common amplitude adjustment trimpot, +5.0 to +100.0 volts to an external 50 ohm load. HV set test point is provided. Output is inherently low impedance so should be terminated in 50 ohms to maintain proper pulse shapes.

Max Duty Cycle 0.3% at 100 volts, 5.0% at 5 volts

Max pulse width xxxx at 100 volts, xxx at 5 volts

Max Duty Cycle 0.3% at 100 volts, 5.0% at 5 volts

Min Pulse Width 8ns

Automatic pulse width limiting prevents transformer saturation

JITTER < 8 ps RMS
CONNECTORS Signals, SMB; optional SMA

Power, 2.5 mm coaxial, center positive

Tip jacks monitors for HV output level

Tip jacks monitors trigger threshold level

INDICATORS Power-on LED
POWER +12 volts, 0. 8 amps max.

Highland model J12 wall-plug supply is recommended.

ENVIRONMENT 0 to 70 degrees C, operating.
0 to 95% humidity, non-condensing.
Highland Technology - Copyright ©2010 All rights reserved
Phone: 415.551-1700  Fax: 415.551-5129
email: info@HighlandTechnology.com